November 2021 : Field Corrective Action Report GLB143074 - Freelite® Human Lambda Free Kit (for use on the SPAPLUS®) : Imprecision at top of calibration curve
Field Corrective Action Report GLB143074 - Freelite® Human Lambda Free Kit (for use on the SPAPLUS®) : Imprecision at top of calibration curve
Affected Medical Device : Freelite® Human Lambda Free Kit (for use on the SPAPLUS®)
MDA Reference Number : MDA/FCA/2021-P170
MDA Registration Number : IVDC8519819-31358
Description :
Imprecision may occur as a result of the decreased separation between calibrator 5 (110.45mg/L) and calibrator 6 (173.76mg/L). When the OD values for calibrator 5 exceed those for calibrator 6 the curve fit may show imprecision which extends down to 70mg/L. Curves exhibiting this fit should not be accepted, as stated in Section D.2.2 of the SPAPLUS Reference Guide (FIN012). Calibration curves must only be accepted if Absorbance (Abs) values increase with increasing concentration (Conc.).