November 2021 : Field Corrective Action Notification From Thermo Scientific Microbiology Sdn. Bhd. - R30163701 Thermo Scientific™ Remel™ Shigella dysenteriae polyvalent, FSN-2021-012 : R30163701 lot 2971158 is not performing to IFU criteria
Field Corrective Action Notification From Thermo Scientific Microbiology Sdn. Bhd. - R30163701 Thermo Scientific™ Remel™ Shigella dysenteriae polyvalent, FSN-2021-012 : R30163701 lot 2971158 is not performing to IFU criteria
Affected Medical Device :
Thermo Scientific™ Remel™ Shigella dysenteriae polyvalent
Model Number: R30163701
Lot Number: 2971158
MDA Reference Number : MDA/FCA/2021-P159
MDA Registration Number :
SA
(8) dlm. MDA.600-3/1/12 Jilid 8
Decription :
An internal technical investigation has determined that R30163701 Shigella dysenteriae polyvalent Lot 2971158 is failing to agglutinate with Sero type 9 and therefore lead to false negative results.