September 2021 : Important Safety Notice VISTA 120 CMS SpO2 Alarm : Software Patch for Vista CMS required
Important Safety Notice VISTA 120 CMS SpO2 Alarm : Software Patch for Vista CMS required
Affected Medical Device : Vista 120 and Central Monitoring Station (CMS) using CMS SW1.64
MDA Reference Number : MDA/FCA/2021-P130
MDA Registration Number : GC3840819-37203
Decription : Draeger has become aware of an event during which an exceeding of SpO2 alarm thresholds triggered an alarm at the bedside monitor Vista 120 as expected but that this did not lead to visual or audible alarming at the Central Monitoring Station (CMS). Numeric values were displayed correctly.