The list below contains Medical Device’s Field Corrective Action for the month of January 2025. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

  • The return of the medical device to the establishment;
  • Modification of the medical device;
  • Exchange of the medical device;
  • Destruction of medical device; or
  • Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

No

Date Received

Title of FCA

Affected Medical Device

MDA Reference Number

MDA Registration Number

Local Establishment Contact Detail

1

02/01/2025

Specificity issues in Post-Mortem Specimens – Liaison® Anti-HBc"Liaison Anti-HBc, product code 310130

LIAISON® ANTI-HBC

MDA/FCA/P1130-14104111-2025

IVDD2249122-109829

DIAGNOSTICARE SDN BHD

fhong@dignosticare.com.my

2

02/01/2025

Access PCT reagent experiencing a high rate of calibration failures with a failure code of “Max Iterate” on Access 2 and UniCel DxI Immunoassay Systems.

ACCESS PCT

MDA/FCA/P1134-18429856-2025

IVDC6748024-166642

BECKMAN COULTER MALAYSIA SDN. BHD.

song01@beckman.com

3

03/01/2025

Pitch Cable Failures on da Vinci X and Xi Tenaculum Forceps and Small GraptorTM

DA VINCI XI SURGICAL SYSTEM, ENDOSCOPIC INSTRUMENT CONTROL SYSTEM, MODEL IS4000

MDA/FCA/P1119-22749179-2024

GC3810021-64842

DTG MEDICAL SDN BHD

regulatory@dtgmedical.com.sg

4

03/01/2025

Grip Cable Failures on da Vinci X and Xi Reusable Instruments with Jaws

DA VINCI XI SURGICAL SYSTEM, ENDOSCOPIC INSTRUMENT CONTROL SYSTEM, MODEL IS4000

MDA/FCA/P1117-41768852-2024

GC3810021-64842

DTG MEDICAL SDN BHD

regulatory@dtgmedical.com.sg

5

09/01/2025

HVAD Controller 2.0 Disable-VAD-Stopped-Alarm (DVSA)FA1455 HVAD Controller 2.0 Disable-VAD-Stopped-Alarm (DVSA)

HEARTWARE® VENTRICULAR ASSIST DEVICE (HVAD®) ACCESSORIES SYSTEM

MDA/FCA/P1136-75285346-2025

GD2555223-122718

MEDTRONIC MALAYSIA SDN BHD

simon.yam@medtronic.com

6

13/01/2025

Updating product-specific Instructions for Use (IFU) for Haemostatic forceps

SURGICAL FORCEPS

MDA/FCA/P1137-62347179-2025

GA826961015518

B. BRAUN MEDICAL INDUSTRIES SDN. BHD.

chee_hong.lam@bbraun.com

7

14/01/2025

Missing IRU Elevator Panel for Alinity m System

ALINITY M SYSTEM

MDA/FCA/P1138-58042219-2025

IVDB7120020-41682

ABBOTT LABORATORIES (MALAYSIA) SDN. BHD.

khenghoe.kwa@abbott.com

8

15/01/2025

Tec 6 Plus vaporizers can deliver reduced concentration of volatile anesthetic agent

TEC 6 PLUS

MDA/FCA/P1139-89761501-2025

GC9756223-128902

GE HEALTHCARE SDN. BHD

premarket.my@ge.com

9

16/01/2025

Risk of finger entrapment during manual movements of the patient tabletop AD7 and AD7X patient tables which are part of Philips Allura and Azurion systems

AZURION 3

MDA/FCA/P1140-89562790-2025

GC99381544218

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

10

16/01/2025

AZURION 5

MDA/FCA/P1141-96278555-2025

GC10785320-48186

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

11

16/01/2025

AZURION 7

MDA/FCA/P1142-86242538-2025


GC82521235717

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

12

16/01/2025

AZURION BIPLANE

MDA/FCA/P1143-51459141-2025

GC59959884718

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

13

16/01/2025

Philips Allura and Azurion systems Risk of patient fall from the table associated with the use of the mattress

AZURION 3

MDA/FCA/P1144-22082165-2025

GC99381544218

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

14

16/01/2025

AZURION 5

MDA/FCA/P1145-93022840-2025

GC10785320-48186

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

15

16/01/2025

AZURION 7

MDA/FCA/P1147-40137943-2025


GC82521235717

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

16

16/01/2025

AZURION BIPLANE

MDA/FCA/P1148-34304365-2025

GC59959884718

PHILIPS MALAYSIA SDN BERHAD

joshua.chan@philips.com

17

18/01/2025

Radiopaque tip of the guide sheath component

GUIDE SHEATH KIT

MDA/FCA/P1150-82029198-2025

GB61352419917

OLYMPUS (MALAYSIA) SDN. BHD.

hideki.nagai@olympus.com

18

20/01/2025

 Reagent Expired (RE) Result Code Potentially Not Produced When Signal Reagent (SR)Pack Is Expired on VITROS® 3600, 5600 and XT 7600 Systems

VITROS IMMUNODIAGNOSTIC PRODUCTS FERTILITY & PREGNANCY HORMONES & PROTEINS

MDA/FCA/P1155-80197836-2025

IVDB7258323-129195

T T MEDICAL MANAGEMENT SDN BHD

ar@ttmedical.com.my