FA1204 9733763 & S9735585 Biopsy Depth Gauge Cycle View Inaccuracy Phase 1

 

 

Affected Medical Device : 

SOFTWARE 9733763 SYNERGY CRANIAL S7
SOFTWARE 9735585 STEALTHSTATION CRANIAL

 

MDA Reference Number : MDA/FCA/2021-P167

 

MDA Registration Number : GD86373862018

 

Description :

Medtronic has identified an anomaly with the StealthStation™ System Cranial software Biopsy Depth Gauge graphical display which presents during navigation in a Cranial Biopsy Procedure. When this anomaly presents, the software enters a state where the Biopsy Depth Gauge is no longer synchronized with the navigational information displayed on the screen. This causes the position of the biopsy needle to be displayed inaccurately.

 