November 2021 : Urgent Medical Device Correction FA1210 : Puritan Bennett™ 980 Series Ventilator Reverse Polarity (C103 Capacitor)
Urgent Medical Device Correction FA1210 : Puritan Bennett™ 980 Series Ventilator Reverse Polarity (C103 Capacitor)
Affected Medical Device : Puritan Bennett™ 980
MDA Reference Number : MDA/FCA/2021-P160
MDA Registration Number : GC95166567018
Decription :
This voluntary medical device correction is being taken following our investigation of six customer reports stating that a total of four ventilators became inoperable during use. Our investigation identified the root cause for these reports to be a manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly.