October 2021 : Field Safety Notice MI506/21/S : Biograph, Biograph Truepoint, Biograph MCT - Improper use of the Patient Handling System (PHS) Pallet/Foot Extension
Field Safety Notice MI506/21/S : Biograph, Biograph Truepoint, Biograph MCT - Improper use of the Patient Handling System (PHS) Pallet/Foot Extension
Affected Medical Device : Biograph, Biograph Truepoint, Biograph MCT
MDA Reference Number : MDA/FCA/2021-P153
MDA Registration Number :
GC28010431417
GC79074213017
Decription :
Potential patient fall hazards could occur when the PHS Pallet/Foot Extension is improperly assembled by the operator. PHS Pallet/Foot Extension patient fall hazards could occur when the extension is improperly used to position the patient head-first. The Pallet/Foot Extension should only be used for positioning the patient’s feet.