Julai 2021 : Preliminary Report FCA 2104 : Automated Blood Coagulation Analyzer CS-1600 - CS-1600/1300 EMC (Electromagnetic Compatibility) compliance issue
Preliminary Report FCA 2104 : Automated Blood Coagulation Analyzer CS-1600 - CS-1600/1300 EMC (Electromagnetic Compatibility) compliance issue
Affected Medical Device : Automated Blood Coagulation Analyzer CS-1600 - CS-1600/1300
MDA Reference Number : MDA/FCA/2021-P104
MDA Registration Number : IVDA3828910416
Decription : An electromagnetic noise (45.8dB) exceeding that of international specification(40.0dB) was observed around 145MHz when the sample probe detects the liquid surface. The noise source is identified as the oscillation output from the operational amplifier mounted on the sample probe liquid detection board PCB No.20026