The list below contains Medical Device’s Field Corrective Action for the month of February 2025. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

  • The return of the medical device to the establishment;
  • Modification of the medical device;
  • Exchange of the medical device;
  • Destruction of medical device; or
  • Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

No.

Date Received

Title of FCA

Affected Medical Device

MDA Reference Number

MDA Registration Number

Local Establishment Contact Detail

1.

03/02/2025

 FA1446 MiniMed and Paradigm Pump Delivery Volume Accuracy (MiniMed 780G System and Accessories)

MINIMED 780G SYSTEM AND ACCESSORIES

MDA/FCA/P1168-72944848-2025

GC4190021-62126

MEDTRONIC MALAYSIA SDN BHD

simon.yam@medtronic.com

2.

04/02/2025

 FA1466 Pipeline Vantage Adverse Braid Behaviors

PIPELINE VANTAGE EMBOLIZATION DEVICE WITH SHIELD TECHNOLOGY

MDA/FCA/P1172-14392084-2025

GD8603721-68288

MEDTRONIC MALAYSIA SDN BHD

simon.yam@medtronic.com

3.

04/02/2025

 FA1466 Pipeline Vantage Adverse Braid Behaviors

PIPELINE VANTAGE EMBOLIZATION DEVICE WITH SHIELD TECHNOLOGY

MDA/FCA/P1173-85796139-2025

GD8603721-68288

MEDTRONIC MALAYSIA SDN BHD

 simon.yam@medtronic.com

4.

05/02/2025

 Hamilton-C2/C3 devices for long-term neonatal ventilation switch to "Ambient State" after 91 days.

MEDICAL VENTILATOR

MDA/FCA/P1174-80588865-2025

GC25545945118

STAR MEDIK SDN BHD

starmedik@gmail.com

5.

06/02/2025

 FIELD CORRECTIVE ACTION, FCA-MUC-2024-001, FORUM 4.4 DICOM Metadata not updated

FORUM

MDA/FCA/P1175-99051464-2025

GB7833715916

CARL ZEISS SDN BHD

siao-yi.won@zeiss.com

6.

06/02/2025

 FA-25003 – Beckman Coulter has identified improper sample motor alignment in certain Access 2 Immunoassay Analyzers, potentially causing irregular sample carousel motion.

ACCESS ULTRASENSITIVE HGH

MDA/FCA/P1176-28009453-2025

IVDB10306623-125414

BECKMAN COULTER MALAYSIA SDN. BHD.

song01@beckman.com

7.

06/02/2025

 FA-001698 – Beckman Coulter was informed by their supplier that lots 1094A and 1095A of ODC0028 RF Latex Calibrator are unsuitable for use with OSR61105 RF Latex reagent for lot numbers above 2541.

RHEUMATOID INFLAMMATORY DISEASE MARKERS

MDA/FCA/P1177-39096301-2025

IVDB8171123-127940

BECKMAN COULTER MALAYSIA SDN. BHD.

song01@beckman.com

8.

12/02/2025

 Getinge Flow-I Sevoflurane Vaporizers: Additional affected units were found.

GETINGE FLOW-I ANESTHESIA SYSTEM

MDA/FCA/P1179-91268165-2025

GC25644700318

IDS MEDICAL SYSTEMS (M) SDN BHD

SriKantLal@idsmed.com

9.

12/02/2025

 Inability to confirm the validity of the Sterile Barrier throughout the product lifecycle for Acrobat-i Positioner Z

ACROBAT-I VACUUM POSITIONER SYSTEM

MDA/FCA/P1181-78391890-2025

GB3824623-138876

RBD HEALTHCARE SDN BHD

sookpei@rbdhealthcare.com

10.

12/02/2025

 Sevoflurane vaporizers: Additional information for all Getinge/Maquet sevoflurane vaporizers

GETINGE FLOW-I ANESTHESIA SYSTEM

MDA/FCA/P1184-64669568-2025

GC25644700318

IDS MEDICAL SYSTEMS (M) SDN BHD

SriKantLal@idsmed.com

11.

12/02/2025

 Sevoflurane vaporizers: Replacement information

GETINGE FLOW-I ANESTHESIA SYSTEM

MDA/FCA/P1185-46407335-2025

GC25644700318

IDS MEDICAL SYSTEMS (M) SDN BHD

SriKantLal@idsmed.com

12.

14/02/2025

 Dade Actin FS Activated PTT Reagent - Control recovery issues with Dade Ci-Trol 2 and Dade Ci-Trol 3

DADE ACTIN FS ACTIVATED PTT REAGENT

MDA/FCA/P1186-26399040-2025

IVDC695822916

SYSMEX (MALAYSIA) SDN. BHD

cheah.kityin@sysmex.com.my

13.

14/02/2025

 URGENT MEDICAL DEVICE CORRECTION Level 1® H-2 Pressure Chamber Door and Latch Replacement Kits

HOTLINE BLOOD AND FLUID WARMER

MDA/FCA/P1189-64294013-2025

GC5311524-185432

PRIMED MEDICAL SDN BHD

primed_07@yahoo.com

14.

14/02/2025

 URGENT MEDICAL DEVICE NOTIFICATION Bivona® Aire-Cuf®, TTS™, Uncuffed, Mid-Range Neonatal/Pediatric Tracheostomy Tube(s) and Bivona Aire-Cuf®, TTS™, Cuffless FlexTend™, TTS™ FlexTend™ Adult Tracheostomy Tube(s)

BIVONA SILICONE TRACHEOSTOMY TUBE

MDA/FCA/P1190-94465734-2025

GC8583224-187732

PRIMED MEDICAL SDN BHD

primed_07@yahoo.com

15.

16/02/2025

 FA-25007 – DxC 500 AU Clinical Chemistry Analyzer: Beckman Coulter found that the system cannot recalculate results when rerun within a calculated test.

DxC 500 AU Clinical Chemistry Analyzer

MDA/FCA/P1191-50540894-2025

IVDA3952124-168118

BECKMAN COULTER MALAYSIA SDN. BHD.

song01@beckman.com

16.

17/02/2025

 Field safety Notice for Hypromellose 0.3% w/v Eye Drops without Preservative (AIN Ophthal Mellose) with safety and performance issues

HYPROMELLOSE 0.3% W/V EYE DROPS WITHOUT PRESERVATIVE

MDA/FCA/P1163-75202480-2025

GB5836723-114988

AIN MEDICARE SDN BHD

ceo@ainmedicare.com.my

17.

17/02/2025

 FA-AM-FEB2025-304: Abbott RealTime HIV-1 assays and the Alinity m HIV-1 assays

ALINITY M HIV-1

MDA/FCA/P1192-40126137-2025

IVDD3748420-39113

ABBOTT LABORATORIES (MALAYSIA) SDN. BHD.

khenghoe.kwa@abbott.com

18.

17/02/2025

 FA-AM-FEB2025-304: Abbott RealTime HIV-1 assays and the Alinity m HIV-1 assays (4N6690)

ABBOTT REALTIME HIV-1 QUALITATIVE

MDA/FCA/P1193-73964962-2025

IVDD8147970817

ABBOTT LABORATORIES (MALAYSIA) SDN. BHD.

khenghoe.kwa@abbott.com

19.

17/02/2025

 FA-AM-FEB2025-304: Abbott RealTime HIV-1 assays and the Alinity m HIV-1 assays (2G3110)

ABBOTT REALTIME HIV-1

MDA/FCA/P1194-58843080-2025

IVDC7467793718

ABBOTT LABORATORIES (MALAYSIA) SDN. BHD.

khenghoe.kwa@abbott.com

20.

17/02/2025

 FA-24066- incorrect expiry date in the 2-dimensional (2D) barcode value assignment sheet for CK-MB Calibrator REF ODR30034, Lot 1059A.

CK-MB

MDA/FCA/P1195-25032649-2025

IVDC6426122-96352

BECKMAN COULTER MALAYSIA SDN. BHD.

song01@beckman.com

21..

17/02/2025

 Updating product-specific Instructions for Use (IFU) for Sterilization Containers

INSTRUMENTS / TOOL FOR REPROCESSING OF MEDICAL DEVICES

MDA/FCA/P1197-84272830-2025

GMD42472682018A

B. BRAUN MEDICAL INDUSTRIES SDN. BHD.

chee_hong.lam@bbraun.com

22.

18/02/2025

 Inability to confirm the validity of the Sterile Barrier throughout the product lifecycle for Acrobat-i Stabilizer Z

ACROBAT-I VACUUM STABILIZER SYSTEM

MDA/FCA/P1182-91407651-2025

GB3805023-133245

RBD HEALTHCARE SDN BHD

sookpei@rbdhealthcare.com

23.

18/02/2025

 FA1432 Neonate Airway Adapter - Unable to be Disconnected from Endotracheal Tube

MICROSTREAM ADVANCE FILTER LINE

MDA/FCA/P1199-93908751-2025

GA2701022-95580

MEDTRONIC MALAYSIA SDN BHD

simon.yam@medtronic.com

24.

19/02/2025

 Assurity and Endurity Pacemaker Inhomogeneous Epoxy Mixing

ASSURITY & ASSURITY MRI PACEMAKERS

MDA/FCA/P1198-55805208-2025

GD71875822118

ABBOTT MEDICAL (MALAYSIA) SDN. BHD.

yumitra.dhorairaja@abbott.com

25.

20/02/2025

 Urgent Field Safety Notice (FSN) - Variation in syringe to line connection

SEDACONDA ACD

MDA/FCA/P1180-82484419-2025

GB9300821-68674

INTEGA SDN. BHD.

lina.chua@integahealth.com