November 2021 : FSCA Notice CAPA-HER-MD-21-002 - Surgical Microscope Leica Microscope M220 F12 : Additional Retention Fixture to secure M220 Optics Carrier
FSCA Notice CAPA-HER-MD-21-002 - Surgical Microscope Leica Microscope M220 F12 : Additional Retention Fixture to secure M220 Optics Carrier
Affected Medical Device : Surgical Microscope Leica M220 F12
MDA Reference Number : MDA/FCA/2021-P168
MDA Registration Number : GMD74389460918A
Description :
The M220 optics carrier may unintentionally drop into the surgical field, risking contact with the patient. The identified defect is a mechanical part failure with the screw that connects the optics carrier to the free end of the swing arm. As a matter of precaution Leica Microsystems has chosen to initiate a field safety corrective action to provide a retention fixture (see Figure 3) to be added on all Leica M220 F12 Surgical Microscopes manufactured before March 9th, 2017 to ensure patient safety and customer satisfaction.