FSCA LASW 21-02 Aptio by Inpeco and FlexLab Automation : Aptio Automation Firmware (FW) for the ADVIA Centaur XP/XPT Interface Module may lead, in specific and rare cases, to an incorrect association of test result(s) to sample ID.

 

 

Affected Medical Device : Interface Module to the ADVIA®Centaur XP/XPT

 

MDA Reference Number : MDA/FCA/2021-P138

 

MDA Registration Number : IVD62653244817A

 

Decription : The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead, in specific and rare cases, to an incorrect association of test result(s) to sample ID. The default configuration of our solution flags any tube with this occurrence with error code 3160 or 0D60 “Sampling Not Successful”, however, an improper management of the sample may bring an incorrect patient result.

 