FSCA ACHC 21-14 ADVIA and Atellica UCFP carryover : Falsely Depressed Enzymatic Hemoglobin A1c Results due to Reagent Carryover from the Urinary/Cerebrospinal Fluid Protein (UCFP) Assay.

 

 

Affected Medical Device : 

ADVIA®1800 Chemistry System
ADVIA®2400 Chemistry System
ADVIA®Chemistry XPT

 

MDA Reference Number : MDA/FCA/2021-P136

 

MDA Registration Number : 

IVDB5744320-51063
IVDB5927720-49887

 

Decription : Siemens Healthcare Diagnostics Inc. has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results and potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_H) results.

 