Ogos 2021 : Initial Report for CR-2021-PD-DXR-001 (FCO 70900050) - CombiDiagnost R90 - Software Package 1.1.1&1.1.2
Initial Report for CR-2021-PD-DXR-001 (FCO 70900050) - CombiDiagnost R90 - Software Package 1.1.1&1.1.2
Affected Medical Device : CombiDiagnost R90
MDA Reference Number : MDA/FCA/2021-P123
MDA Registration Number : GC25585319317
Decription : After the user selects the Lock-in command, the kV and mA values are not locked immediately and can change if the collimator shutter position is changed directly after the Lock-in command. This will lead to a change of technique factors. If the user stops and restarts fluoroscopy then the Lock-in works as specified and collimation can be changed to the needs of the examination.