Julai 2021 : Field Corrective Action Notification 200-06-103-090 : Digital Accelerator - CITB external interlock check
Field Corrective Action Notification 200-06-103-090 : Digital Accelerator - CITB external interlock check
Affected Medical Device :
Elekta Synergy
Precise Digital Accelerator
Elekta Infinity
Versa HD
MDA Reference Number : MDA/FCA/2021-P106
MDA Registration Number :
GC73830983018
GC227501186818
GC23765696518
GC76319499317
Decription : Check the connections to the CITB for the customer required external interlocks and confirm with the customer that they are correct. Update the calibration values to enable the External Terminate and Room Doors 3 inputs (software interlocks) which have been disabled by default in the calibration.