FIELD CORRECTIVE ACTION LISTING FOR DECEMBER 2023
The list below contains Medical Device’s Field Corrective Action for the month of December 2023. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.
Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –
- The return of the medical device to the establishment;
- Modification of the medical device;
- Exchange of the medical device;
- Destruction of medical device; or
- Specific advice on the use of the medical device.
If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.
No |
Date received |
Title of FCA |
Affected Medical Device |
MDA Reference Number |
MDA Registration Number |
Description |
Local Establishment Contact Detail |
1 |
12 Dec 2023 |
ALERE ACTIM PARTUS |
MDA/FCA/P0612-36284740-2023 |
IVDB16716341018 |
Recent interference testing has shown that the presence of lubricants in the vaginal specimen may interfere with the Partus test result. The possible interference has not been stated in current instructions for use. |
WINNIE HII LIN LIN SENIOR REGULATORY AFFAIRS SPECIALIST ABBOTT DIAGNOSTICS HEALTH SDN BHD |
|
2 |
19 Dec 2023 |
ACTIM® PROM TEST |
MDA/FCA/P0629-63815712-2023 |
IVDB48023332118 |
Our recent interference testing has shown that the presence of lubricants in the vaginal specimen may interfere with the PROM test result, which deviates from what is presented in instructions for use. |
WINNIE HII LIN LIN SENIOR REGULATORY AFFAIRS SPECIALIST ABBOTT DIAGNOSTICS HEALTH SDN BHD |
|
3 |
29 Dec 2023 |
CARESCAPE R860 |
MDA/FCA/P0642-75016512-2023 |
GC97483321017 |
GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engström Carestation/ Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40C (104F), at the lowest possible flow condition of 2 L/min (worst-case, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). |
LIM SOCK SIANG DIRECTOR GE HEALTHCARE SDN. BHD |
|
4 |
23 Dec 2023 |
ELECTROSURGICAL SYSTEM |
MDA/FCA/P0614-42120567-2023 |
GC2965522-109913 |
The purpose of this medical device notification is to communicate an important change to the intended use population (e.g., indications) for the MEGA SOFT™ Universal and Universal Plus Reusable Patient Return Electrodes (product codes: 0845, 0846, 0847, and 0848) to help ensure safe and effective use. |
YAP KWAI LING REGULATORY AFFAIRS LEAD, MEDTECH JOHNSON & JOHNSON SDN BHD |
|
5 |
21 Dec 2023 |
HEARTWARE VENTRICULAR ASSIST SYSTEM (HVAD) |
MDA/FCA/P0636-11042593-2023 |
GD74345985518 |
CAPA 502194 was initiated on 23-Jul-2020 due to the escalation of RPI IEDF00247. RPI IEDF00247’s scope includes complaints received on pumps failing to initially start, restart or experiencing a delay in restart within the maximum number of restart attempts per the software configuration (30 restart attempts), regardless of whether the pump ultimately restarted. |
CHOW SOON YEE REGULATORY AFFAIRS LEAD MEDTRONIC MALAYSIA SDN BHD |
|
6 |
22 Dec 2023 |
STEALTHSTATION S7 TREATMENT GUIDANCE SYSTEM (CRANIAL) |
MDA/FCA/P0641-26011160-2023 |
GD5865623-131796 |
his FCA for software update of new Cranial Software (9735585), Version 3.1.5 to correct the potential for inaccuracy during biopsy procedures using the StealthStation™ S7 and i7 Biopsy Depth Gauge feature. |
CHOW SOON YEE REGULATORY AFFAIRS LEAD MEDTRONIC MALAYSIA SDN BHD |
|
7 |
20 Dec 2023 |
EVIS EXERA III DUODENOVIDEOSCOPE TJF-Q190V |
MDA/FCA/P0617-65322625-2023 |
GB4632321-70352 |
This Field Safety Notice is initiated after becoming aware that the Single Use Distal Cover MAJ-2315 may unexpectedly detach from the duodenoscope during use. |
YEOH SEO CHING, SHELLY MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS OLYMPUS (MALAYSIA) SDN. BHD. |
|
8 |
15 Dec 2023 |
AZURION 3 |
MDA/FCA/P0617-80515884-2023 |
GC99381544218 |
If this issue occurs, the Philips Azurion system might not recover from this fault situation causing the system to not be available. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure. |
ZAM ZARINA BINTI MAT AIL QUALITY & REGULATORY MANAGER PHILIPS MALAYSIA SDN BERHAD |
|
9 |
18 Dec 2023 |
N ANTISERUM TO HUMAN IGA |
MDA/FCA/P0627-97405132-2023 |
IVDC3972923-114643 |
Siemens Healthcare Diagnostics Products GmbH has confirmed that the Instruction for Use (IFU) for N Antiserum to Human IgA and for N Antiserum to Human IgM might be misleading concerning that the reagents show no high-dose hook effect up to 58.6 g/L and 44.6 g/L respectively for the sensitive applications (IgAs, SAN 22 and IgMs, SAN 23). |
LAM YEN LI HEAD OF QUALITY & TECHNOLOGY SIEMENS HEALTHCARE SDN. BHD. |
|
10 |
18 Dec 2023 |
N ANTISERUM TO HUMAN IGM |
MDA/FCA/P0628-19341860-2023 |
IVDC3764423-118238 |
Siemens Healthcare Diagnostics Products GmbH has confirmed that the Instruction for Use (IFU) for N Antiserum to Human IgA and for N Antiserum to Human IgM might be misleading concerning that the reagents show no high-dose hook effect up to 58.6 g/L and 44.6 g/L respectively for the sensitive applications (IgAs, SAN 22 and IgMs, SAN 23). |
LAM YEN LI HEAD OF QUALITY & TECHNOLOGY SIEMENS HEALTHCARE SDN. BHD. |