The list below contains Medical Device’s Field Corrective Action for the month of October 2023. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

(a)     The return of the medical device to the establishment;

(b)    Modification of the medical device;

(c)     Exchange of the medical device;

(d)    Destruction of medical device; or

(e)    Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

 

No

Date received

Title of FCA

Affected

Medical Device

MDA Reference Number

MDA Registration Number

Description

Local Establishment

Contact Detail

1

8 November 2023

Urgent Field Safety Notice on Software Upgrade of ID NOW Instrument

ID NOW™ INSTRUMENT

MDA/FCA/P0564-22155588-2023

IVDC4177623-150835

When using the current version of ID NOW software, version 7.0, users have the ability to run the ID NOW COVID-19 2.0 and ID NOW Influenza A&B 2 test sequentially, from one patient sample. Some customers have reported an increase in Influenza B false positive test results when using the device in this manner.

WINNIE HII LIN LIN

SENIOR REGULATORY AFFAIRS SPECIALIST

ABBOTT DIAGNOSTICS HEALTH SDN BHD

2

8 November 2023

Exchange of Transfer Set with possible leaking issue

TRANSFER SET

MDA/FCA/PX0587-75794058-2023

GA84078400417

Complaints received on "Transfer Set" leaking which indicates tubing failure in tubing

NOOR IBTISYAM BINTI MOHAMAD ADNAN

REGULATORY AFFAIRS EXECUTIVE

ADVENTA HEALTHCARE SDN. BHD.

3

30 October 2023

FSCA-2023-08-14 (1) // Infusomat® Space - Upstream Occlusion Sensor

INFUSION PUMP

MDA/FCA/PX0540-92348718-2023

GC92724141717

In the course of Post Market Surveillance activities, we identified a sporadic occurrence of potentially false downand upstream pressure alarms which are caused by the upstream occlusion pressure sensor of the Infusomat® Space Infusion Pump.

QUEK LANG FAH

HEAD OF REGULATORY AFFAIRS

B. BRAUN MEDICAL INDUSTRIES SDN. BHD.

4

6 November 2023

Field Corrective Action for IndiGo Drive Assistance for Arjo Enterprise 5000x, Enterprise 8000X, Enterprise 9000X, Citadel medical beds

ENTERPRISE 9000X

MDA/FCA/P0550-24471944-2023

GMD85545422818A

Malfunction leading to the unintended movement of bed wheels

CHONG WAI MUN

GENERAL MANAGER

BEST CONTACT (M) SDN BHD

5

4 October 2023

MA 2023 0006 DRX Revolution Mobile X-ray System CPI Generator Issue

DRX-REVOLUTION MOBILE X-RAY SYSTEM

MDA/FCA/P0530-99906041-2023

GC52334161417

There is a potential for unexpected failure of the electrical components within the Communication & Power Industries, Inc (CPI) generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self-contained thermal overload within the generator.

WNG AH LIN

BUSINESS ADMINISTRATOR

CARESTREAM HEALTH (M) SDN BHD

6

6 October 2023

Revision on stability sections in Product Insert- Lyphochek Hemoglobin A2 Control

HEMOGLOBIN CONTROL

MDA/FCA/P0544-63864384-2023

IVDB3773471217

Bio-Rad, the manufacturer has observed that Hemoglobin F may not meet the reconstituted refrigerated (2 to 8°C) storage stability claims as printed in the package insert. The stability claims for Hemoglobin F have been revised as indicated by the highlighted text and bold font in the FSN.

ADELINE NG POOI KUAN

GROUP CORPORATE COMPLIANCE SENIOR MANAGER

CHEMOPHARM SDN BHD

7

22 November 2023

Possible shutdown of Dräger anesthesia workstation Atlan due to possible backup battery failures

ATLAN

MDA/FCA/P0597-86266273-2023

GC7299919-37202

Within our market surveillance activities regarding our anesthesia workstation Atlan we became aware of a few cases in which the internal backup battery failed spontaneously while the Atlan is being operated without mains supply. This resulted in an unexpected shutdown of the device while it was running on batteries.

BEATRICE MAH WAN XIN

QUALITY AND REGULATORY AFFAIR SPECIALIST

DRAEGER MALAYSIA SDN BHD

8

17 October 2023

Potential for Loose Universal Surgical Manipulators (USM) on da Vinci X and Xi Systems

DA VINCI XI SURGICAL SYSTEM, ENDOSCOPIC INSTRUMENT CONTROL SYSTEM, MODEL IS4000

MDA/FCA/P0553-57377033-2023

GC3810021-64842

A population of the da Vinci X and Xi system instrument arms (USMs) have Instrument Carriages that may separate from the Insertion Axis Linear Rail during shipping, handling, or repositioning of the arms when the system is powered down. The Instrument Carriage would remain attached to the drive motor which prevents free-fall motion or complete detachment from the USM

RYAN YEO CHOON HWEE

REGULATORY AFFAIRS MANAGER, SOUTH EAST ASIA

DTG MEDICAL SDN BHD

9

26 October 2023

Flushing Difficulties DPT Occlusion Issue

TRUWAVE DISPOSABLE PRESSURE TRANSDUCER AND VAMP SYSTEM

MDA/FCA/P0562-24787716-2023

GB838093516

As per product evaluation, the Flow restrictor was located further into the housing and blocked the flow path below the poppet. Therefore, Acumen IQ units were not able to be primed due to occlusion within flush device.

WONG WAI YI

REGULATORY AFFAIRS SPECIALIST

EDWARDS LIFESCIENCES (MALAYSIA) SDN BHD

10

24 October 2023

Flushing Difficulties DPT Occlusion Issue

HEMOSPHERE ADVANCED MONITORING PLATFORM

MDA/FCA/P0563-39428819-2023

GC766371236819

11

23 October 2023

  Stay safe Disinfection Cap (article no. 2845091, F00005376, F00005996) – potential side effects of hypothyroidism and hypersensitivity

PD TUBING SYSTEM

MDA/FCA/P0531-92558309-2023

GB5396719-32241

Based on current literature, it has been recognized that an increased exposure to povidone-iodine, a potent antiseptic present in the stay safe Disinfection Cap used to close the PD catheter extension could lead to iodine-induced hypothyroidism and associated clinical sequelae.

SHARON DING WENSZE

REGULATORY AFFAIRS MANAGER

FRESENIUS MEDICAL CARE MALAYSIA SDN BHD

12

10 October 2023

Flexiview 8800, OEC 9800, OEC 9900 and OEC Elite systems with 9-inch Image Intensifier - systems can become detached if the systems encounter a large impact force when moved.

OEC ELITE

MDA/FCA/P0547-33670176-2023

GC35179812618

GE HealthCare has become aware that the 9-inch Image Intensifier for OEC Flexiview 8800, OEC 9800, OEC 9900 and OEC Elite systems can become detached if the systems encounter a large impact force when moved. No instances of this issue have occurred in clinical use. GE HealthCare became aware of one instance of image intensifier detachment during shipment.

LIM SOCK SIANG

DIRECTOR

GE HEALTHCARE SDN. BHD

13

10 October 2023

Flexiview 8800, OEC 9800, OEC 9900 and OEC Elite systems with 9-inch Image Intensifier - systems can become detached if the systems encounter a large impact force when moved.

OEC ELITE

MDA/FCA/P0548-40580869-2023

GC6766523-114390

14

12 October 2023

Field Corrective Action for ExtraFit Latex Examination Gloves, Powdered with Powder Issue

LATEX EXAMINATION GLOVES, POWDERED

MDA/FCA/PX0520-35151668-2023

GA3428235816

Meditech Gloves Sdn. Bhd. has initiated a field corrective action for the ExtraFit Latex Examination Gloves, Powdered product due to customer complaints received, indicating issues with uneven powder and clumps that have transferred to both equipment and patients who have come into contact with the gloves despite having powder content well within the specification limit.

HASNAH BINTI ABDUL HAMID

QUALITY ASSURANCE MANAGER

MEDITECH GLOVES SDN BHD

15

2 October 2023

StealthStation™ S8 App version 2.0 and 2.0.1 StealthStation™ S8 and StealthStation FlexENT™ Software Plan Data Shift with New Reference Exam

STEALTHSTATION S8 CRANIAL SYSTEM

MDA/FCA/P0535-60608026-2023

GC7686023-140750

Medtronic has identified a software anomaly in StealthStation™ S8 Software App versions 2.0 and 2.0.1 that can occur within a Cranial (including DBS and Stereotaxy) or ENT procedure type, while merging exams with StealthMerge™ or StealthMerge™ ENT software,

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

16

STEALTHSTATION S8 SPINAL SYSTEM

MDA/FCA/P0536-34928297-2023

GC6531123-143275

17

STEALTHSTATION S8 ENT SYSTEM

MDA/FCA/P0537-60192016-2023

GC771281137418

18

STEALTHSTATION FLEXENT SYSTEM

MDA/FCA/P0538-55405759-2023

GC6421521-58428

19

10 October 2023

FA1370 McGRATH MAC Video Laryngoscopes Misappropriated Devices

McGRATH MAC Video Laryngoscope

MDA/FCA/P0542-68105454-2023

GA7934120-46549

A number of McGrath MAC Video Laryngoscopes [SKU 301-000-000] dispositioned as scrap at the contract manufacturer (CM) Plexus has been found for sale on social media platforms, Facebook Market Place and Mercado Libre.

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

20

5 October 2023

FA1255 NIM Endotracheal Tube Airway Blockage Issue

NIM-NEURO 3.0 SYSTEM AND ACCESSORIES

MDA/FCA/P0543-40274208-2023

GB45802845818

Medtronic has received complaint reports from customers experiencing issues with the NIM™ Standard Reinforced EMG Endotracheal Tube & NIM CONTACT™ Reinforced EMG Endotracheal Tube during use. It was reported that the cuff of the endotracheal EMG tube was over-inflated, causing it to herniate over the end of the tube leading to loss of ventilation due to airway blockage.

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

21

13 October 2023

FA1325 Minimed Mio Infusion Incorrect Malay IFU

MINIMED MIO INFUSION SET

MDA/FCA/P0551-21039693-2023

GC8179022-89707

Every home use device marketed by Medtronic in Malaysia must contain the Instructions for Use (IFU) with correct information in the Malay language. Per the current process at Medtronic Malaysia, when material is received, the packaging box is reworked to include the Malay language IFU

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

22

20 October 2023

FA1249 Broken Battery Cap Contact Issue

MINIMED 780G SYSTEM AND ACCESSORIES

MDA/FCA/P0555-27176012-2023

GC4190021-62126

A reliability analysis was conducted and determined that all consignees with pumps aged older than 1 year should receive backup battery caps. The reliability report noted that the battery cap metal contact issue is a wear out phenomenon and therefore the risk of the pump losing power increases the longer the pump is in use.

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

23

MINIMED 700 INSULIN PUMP AND KIT

MDA/FCA/P0556-30139830-2023

GC8922821-64034

24

MINIMED 640G SYSTEM AND ACCESSORIES

MDA/FCA/P0557-96661538-2023

GC8430922-106837

25

27 October 2023

FA1372 Rev_MCS Useful Life IFU Update

HEARTWARE VENTRICULAR ASSIST SYSTEM (HVAD)

MDA/FCA/P0572-51089984-2023

GD74345985518

The HVAD System has features and components, such as the internal battery in the controller, which are expected to eventually wear out or otherwise reach the end of their useful life, as is inherent in any such physical system. To help inform on how long the controller could be expected to perform before reaching end of useful life, the existing Instructions for Use and Patient Manual contain training and guidance (see Table 5) on length of useful life and management of the controller.

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

26

16 October 2023

FA1325 Minimed Mio Infusion Incorrect Malay IFU

MINIMED MIO INFUSION SET

MDA/FCA/PX0552-67361328-2023

GC8179022-89707

Every home use device marketed by Medtronic in Malaysia must contain the Instructions for Use (IFU) with correct information in the Malay language. Per the current process at Medtronic Malaysia, when material is received, the packaging box is reworked to include the Malay language IFU.

CHOW SOON YEE

REGULATORY AFFAIRS LEAD

MEDTRONIC MALAYSIA SDN BHD

27

30 October 2023

Potential for uncontrolled extension of XVI / iViewGT™ detector arms

ELEKTA MEDICAL LINEAR ACCELERATOR (EMLA)

MDA/FCA/P0574-45275425-2023

GC8284822-112386

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT™ / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.

BHARAT KUMAR A/L VRAJLAL PREMCHAND

DIRECTOR

NYPRAX BUSINESS SOLUTIONS

28

2 October 2023

EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190 do not start up properly when the power is turned on.

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE SYSTEM

MDA/FCA/P0515-24057168-2023

GB41963114717

Olympus found that certain lots of the CV-190 do not start up properly and as a result, the image from the endoscope is not displayed when the operations listed below are performed because parts that deviated from the specification were assembled into the power supply unit.

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD.

29

4 October 2023

Amendment to the Instruction for Use

HIGH FLOW INSUFFLATION UNIT

MDA/FCA/P0528-92249188-2023

GB929511113418

The overpressure condition was caused by the following two Hazardous Situations.

(1) Overpressure without device abnormality (UHI-4 normal state without abnormality)

(2) Over pressurization due to uncontrolled cavity pressure caused by equipment failure. (UHI-4 failure state)

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD.

30

2 October 2023

Revised Instructions of Updated Drying Procedure for the GIF-1TH190.

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE SYSTEM

MDA/FCA/P0534-71896576-2023

GB41963114717

This Field Safety Notice is based on the reprocessing validation for a Catch-Up 510(k) application to be submitted to the FDA on this model. The reprocessing validation results for EOG sterilization of the GIF-1TH190 did not meet the criteria of “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff”, issued March 17, 2015 (2015 reprocessing guidance).

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD.

31

17 October 2023

Recommendation to utilize the Lock touch panel function and follow the instruction manual including the WARNING.

EVIS X1 VIDEO SYSTEM CENTER

MDA/FCA/P0549-18524426-2023

GB2460720-48587

This Field Safety Notice based on the complaint received outside Malaysia that the water delivery tube hit the touch panel and the air delivery button was unintentionally turned on during the procedure.

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD.

32

26 October 2023

Reminding the WARNINGS within Olympus bronchoscope’s Operation Manual.

EVIS LUCERA ELITE BRONCHOVIDEOSCOPE SYSTEM

MDA/FCA/P0567-89902217-2023

GB66358142517

This Field Safety Notice based on the complaint received outside Malaysia that the water delivery tube hit the touch panel and the air delivery button was unintentionally turned on during the procedure.

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD.

33

EVIS EXERA III BRONCHOVIDEOSCOPE SYSTEM

MDA/FCA/P0569-47659304-2023

GB58878308917

34

OPTERA VIDEO SYSTEM

MDA/FCA/P0571-87040392-2023

GB15542818418

35

2 November 2023

Flexible Endoscopes Serviced at Olympus Repair Centre May Have Residual Channel Moisture.

EVIS EXERA III BRONCHOVIDEOSCOPE SYSTEM

MDA/FCA/P0579-93618134-2023

GB58878308917

This Field Safety Notice is based on the 3 complaints received outside Malaysia about water dripping off from the distal end of demo/loaner endoscopes. According to the Inspection Manual issued by OMSC (Olympus Medical Systems Co., Ltd), repair production control had instructed the worldwide repair centers to drain water from the entire channel system after inspection and check if there is no water residue in the channel.

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD.

36

7 November 2023

Reminding to inspect the ENDOEYE HD II image & have spare as described in the Instructions For Use (IFU).

ENDOEYE HD II

MDA/FCA/P0584-16111136-2023

GB45912709718

This Field Safety Notice is initiated based on an increased complaint rate outside Malaysia that is related to “pink or green images” of the Video Laparoscopes ENDOEYE HD II (WA50040A and WA50042A).

YEOH SEO CHING, SHELLY

MANAGER, QUALITY ASSURANCE & REGULATORY AFFAIRS

OLYMPUS (MALAYSIA) SDN. BHD

37

5 October 2023

 

2023-CC-HPM-022

Xper Flex Cardio Device, Model #: FC2010 & FC2020, used with the Xper Flex Cardio Physiomonitoring System and Philips Hemo System

 

XPER FLEX CARDIO PHYSIOMONITORING SYSTEM

MDA/FCA/P0541-81082167-2023

GC95469295617

Following exposure to defibrillation voltage, the system returns to normal operating mode within 5 seconds. However, if the Xper Flex Cardio Devicehas been exposed to greater than the specified energy related to the defibrillator and/or ESU use, it might fail to return to normal operating mode within 5 seconds and it may require a restart in order to continue ECG monitoring.

ZAM ZARINA BINTI MAT AIL

Q&R MANAGER

PHILIPS MALAYSIA SDN BERHAD

38

5 October 2023

2023-IGT-BST-002 - Philips Azurion System R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 Potential Loss of X-ray Functionality

AZURION 5

MDA/FCA/P0545-22716940-2023

GC10785320-48186

Philips has identified a potential safety issue where the Philips Azurion system may unexpectedly lose X-ray functionality. Due to a software issue, a mechanism that is present in the system to manage the number and size of Log Trace Files does not function properly.

ZAM ZARINA BINTI MAT AIL

Q&R MANAGER

PHILIPS MALAYSIA SDN BERHAD

39

23 October 2023

 

FSN-2023-CC-EC-026

HeartStart Intrepid Monitor/Defibrillator (867172) Service Manual Electrical Safety Test Issue

 

HEARTSTART INTREPID

MDA/FCA/P0561-32186601-2023

GC2897720-42451

The current HeartStart Intrepid Service Manual does not detail IEC Class I electrical safety test verification methods. The lack of this testing does not introduce a failure, but it may prevent a failure from being detected. This could occur when an IEC Class I device is tested as a Class II device after a servicing event.

ZAM ZARINA BINTI MAT AIL

Q&R MANAGER

PHILIPS MALAYSIA SDN BERHAD

40

26 October 2023

 

FSN 2023-PD-MR-014 Achieva 1.5T,

Gradient Coil component failure may act as a heat source with potential to produce smoke and/or fire.

 

ACHIEVA SERIES

MDA/FCA/P0566-89461896-2023

GB44884842218

hilips has identified an issue where a specific component failure in the Gradient Coil of the affected 1.5T MR Systems listed in this letter may act as a heat source with a potential to produce smoke and/or fire.

ZAM ZARINA BINTI MAT AIL

Q&R MANAGER

PHILIPS MALAYSIA SDN BERHAD

41

16 November 2023

FSN-2022-IGT-BST-002 L-arm rotation cover may fall on Philips Allura with monoplane fixed ceiling mounts.

ALLURA XPER FD SERIES

MDA/FCA/P0588-48405519-2023

GC59365128817

Philips has identified a potential safety issue with the Philips Allura and Azurion Product Families having monoplane fixed ceiling mounted systems. The ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur.

ZAM ZARINA BINTI MAT AIL

Q&R MANAGER

PHILIPS MALAYSIA SDN BERHAD

42

AZURION 7

MDA/FCA/P0589-90484518-2023

GC82521235717

43

4 October 2023

Unsecure blade of ACROBAT-I Vacuum Stabilizer System

ACROBAT-I VACUUM STABILIZER SYSTEM

MDA/FCA/P0522-28049082-2023

GB3805023-133245

Maquet/Getinge is initiating a voluntary Medical Device Field Safety Notice-Removal for the ACCESSRAIL Platform Standard Blade and Deep Blade due to the blade not securely latching onto the Activator Drive.

HEW SOOK PEI

ADMIN MANAGER

RBD HEALTHCARE SDN BHD

44

31 October 2023

Unintended Movement from travel unit at MEERA operating tables

MEERA, MEERA ST

MDA/FCA/P0575-35899350-2023

GA2599721-82114

Under certain conditions, we have identified that an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec.

HEW SOOK PEI

ADMIN MANAGER

RBD HEALTHCARE SDN BHD

45

20 November 2023

ResMed Masks with Magnets and Potential Magnetic Interference with Certain Medical Devices

AIRFIT N20

MDA/FCA/P0591-73546052-2023

GB95390436717

The reported injuries included pacemaker failure leading to pacemaker replacement, a need for shunt adjustment, resetting of automatic implantable cardioverter defibrillator, arrhythmia, cognitive changes, headaches, change in heartrate (tachycardia, bradycardia), convulsions (seizures), and irregular blood pressure.”

NAZATUL ASHIKIN BINTI REZUAN

SENIOR QUALITY ASSURANCE ENGINEER

RESMED MALAYSIA OPERATIONS SDN. BHD.

46

11 October 2023

 

Surgical Microscopes: M530 OHX, PROvido, Proveo 8

FSCA Identifier: CAPA-LIS-MD-22-001 Replacement of Power Supply Unit

 

PROVIDO Surgical Microscope

MDA/FCA/P0546-90235430-2023

GA6844423-140492

Leica Microsystems has become aware of process weaknesses of the sub-supplier producing the affected Power Supply Unit. These process weaknesses have caused that the Power Supply can become overheated due to internal component overstressing.

NOOR FAIZAH BINTI KAMARUDIN

REGULATORY AFFAIRS EXECUTIVE

SCHMIDT BIOMEDTECH SDN BHD

47

21 November 2023

fabian HFO, fabian +nCPAP evolution, and fabian Therapy evolution Multiple Issues related to Software Anomalies for Field Safety Corrective Action FSCA-21-002

ABIAN NEONATAL / PEDIATRIC VENTILATORS

MDA/FCA/P0594-37401721-2023

GC92488887018

The purpose of this communication is to inform you that a software upgrade (software version 5.2.1) and revised Instructions for Use associated with FSCA-21-002, as communicated in the original Field Safety Notice (FSN) FSCA-21-002_FSCA-21-003-FSN-1, are now available for the fabian HFO, fabian +nCPAP evolution, and fabian Therapy evolution ventilators.

NOOR FAIZAH BINTI KAMARUDIN

REGULATORY AFFAIRS EXECUTIVE

SCHMIDT BIOMEDTECH SDN BHD

48

5 October 2023

053 FSCA POC 23-018 RAPIDPoint 500/500e Perhexiline Maleate and Atomoxetine Hydrochloride Interference on Sodium Measurement

RAPIDPOINT 500

MDA/FCA/P0539-38330352-2023

IVDC6984796818

Siemens Healthcare Diagnostics Inc. has determined that two drugs, Perhexiline Maleate and Atomoxetine Hydrochloride, may interfere with Sodium results that are reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems.

LAM YEN LI

HEAD OF QUALITY & TECHNOLOGY

SIEMENS HEALTHCARE SDN. BHD.

49

26 October 2023

001 FSCA POC 24-001 epoc Blood Analysis System - Epoc NXS Host Potential for Test Record to Include Unselected Analytes

EPOC BLOOD ANALYSIS SYSTEM

MDA/FCA/P0560-73047662-2023

IVDC7557621-75750

Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under specific conditions when the (automatic) Delete Blood Tests feature is enabled.

LAM YEN LI

HEAD OF QUALITY & TECHNOLOGY

SIEMENS HEALTHCARE SDN. BHD.

50

26 October 2023

002 CN AIMC 24-01 Atellica IM, Atellica CI, and ADVIA Centaur Free Prostate-Specific Antigen (fPSA) Invalid Calibrations

ATELLICA® IM FREE PROSTATE‑SPECIFIC ANTIGEN (FPSA)

MDA/FCA/P0570-60011856-2023

IVDC82776266818

Our records indicate that your facility may have received the product(s) listed in Table 1. Siemens Healthcare Diagnostics Inc. has confirmed the potential for invalid calibrations with the fPSA assay due to the “low calibrator deviation” parameter resulting outside of the defined specification range.

LAM YEN LI

HEAD OF QUALITY & TECHNOLOGY

SIEMENS HEALTHCARE SDN. BHD.

51

31 October 2023

003 CN POC 24-003 epoc Blood Gas Analysis System - Epoc Host2 Delay of Results and Potential Loss of Patient Sample When Using the Positive Patient ID (PPID) Lookup

EPOC BLOOD ANALYSIS SYSTEM

MDA/FCA/P0576-26085425-2023

IVDC7557621-75750

Siemens Healthcare Diagnostics Inc. has confirmed an occurrence of the epoc Host2 device freezing during analysis when the “Use this ID” button is pressed while the system is performing test result calculations. This issue only occurs when the epoc Host2 is used with the PPID feature during calculations.

LAM YEN LI

HEAD OF QUALITY & TECHNOLOGY

SIEMENS HEALTHCARE SDN. BHD.

52

7 November 2023

006 FSCA POC 24-004 RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) - Potential for Low Na+ Values and Question Result Error flags 10_53.pdf

RAPIDPOINT 500

MDA/FCA/P0583-84567939-2023

IVDC6984796818

Complaints were received regarding discrepant Na+ results (low Na+ Bias). The associated complaints are captured in the Parent investigation ticket 2023-00013922. The investigation of this issue focuses on the Na+ performance given that some of the alleged biases exceed the total allowable error of > +/- 4mM (1mM = 1mmol/L).

LAM YEN LI

HEAD OF QUALITY & TECHNOLOGY

SIEMENS HEALTHCARE SDN. BHD.

53

21 November 2023

007 CN DC 24-01 Dimension & Dimension Vista Iron Flex reagent - Dimension and Dimension Vista IRON Lipemia Interference Updated in the Instructions for Use

DIMENSION® IRON FLEX® REAGENT CARTRIDGE

MDA/FCA/P0596-98021662-2023

IVDB3764328117

Siemens Healthineers’ internal testing has determined that the IRON assay, for use on the Dimension Clinical Chemistry System, has lipemia interference claims in the Instructions For Use (IFU) that require updates as shown in Table 2. While Siemens has not received any customer complaints for this issue, grossly lipemic samples have the potential to falsely elevate iron results on the Dimension Clinical Chemistry System.

LAM YEN LI

HEAD OF QUALITY & TECHNOLOGY

SIEMENS HEALTHCARE SDN. BHD.

54

2 November 2023

Gem PAKs for Gem Premier 3000/3500, 4000 with iQM

GEM® PREMIER™ 3500 SYSTEM

MDA/FCA/P0577-83510019-2023

IVDC7302023-115078

This notification is intended to advise you regarding a potential performance issue with the following specific GEM PAKs for the GEM Premier 3000/3500, 4000 with iQM/iQM2 (Intelligent Quality Management)

CHAI AI LIN

MANAGER

STRAITS SCIENTIFIC (M) SDN BHD

55

2 November 2023

GEM PREMIER 5000 BLOOD GAS SYSTEM

MDA/FCA/P0578-38696993-2023

IVDC33500201118

56

30 October 2023

URGENT PRODUCT CORRECTION NOTIFICATION Potential for Incorrect Rh Results on the ORTHO VISION® Analysers and Ortho Optix™ Reader

ORTHO VISION ANALYZER SYSTEM

MDA/FCA/P0573-38882849-2023

IVDD4853923-138156

During an internal review, QuidelOrtho identified that Test ID 10023 included calculated Rh (Anti-D or RhD) Interpretation Result when in fact, no Anti-D column was used for the test.

MENG TSE LEEAU

MANAGING DIRECTOR

T T MEDICAL MANAGEMENT SDN BHD