November 2021 : Initial Report for CR-2021-IGT-BST-011 (FCO72200484) - Leakage of cooling fluid of flat detector in the R-cabinet ®
Initial Report for CR-2021-IGT-BST-011 (FCO72200484) - Leakage of cooling fluid of flat detector in the R-cabinet ®
Affected Medical Device : Philips Allura Xper
MDA Reference Number : MDA/FCA/2021-P173
MDA Registration Number : GC97115121117
Description :
Philips has discovered a problem with the flat detector cooling system. Due to a leak in the flat detector cooling system, cooling liquid may leak outside the drip tray of the Laird chiller. If this occurs, the liquid may drip into electrical components in the R-cabinet located in the technical room which could cause damage to the system.